评定连续流动法测量烟用接装纸中六价铬的测量不确定度
Determination Uncertainty of Hexavalent Chromium of Cigarette Paper by Continuous Flow Method
DOI: 10.12677/AAC.2017.72014, PDF, HTML, XML, 下载: 1,545  浏览: 3,905 
作者: 李远根, 王 龙, 寇天舒:河北中烟工业有限责任公司,河北 石家庄
关键词: 连续流动法烟用接装纸六价铬不确定度Continuous Flow Method Cigarette Paper Hexavalent Chromium Uncertainty Assessment
摘要: 为评定连续流动法烟用接装纸中六价铬的测量不确定度,本实验分析了测量过程,修订了连续流动法烟用接装纸中六价铬的测定因果关系图,建立了测量不确定度评定的数学模型。通过对测量过程中各环节的评定,得出连续流动法烟用接装纸中六价铬测量结果的不确定度的主要来源为:样品的含水率,连续流动分析仪测定的样品浓度和测量重复性。测量结果的标准不确定度为0.12%,在95%的置信度下,取包含因子为k = 2,其扩展不确定度为0.24%。
Abstract: In order to evaluate uncertainty in hexavalent chromium of cigarette paper, a new math stereotype was set up to describe the whole process. The model was derived from analysis of hexavalent chromium measurement of cigarette paper and modified causation figure. The water content, ex-press value and repeatability were considered to be the main sources of uncertainty in hexavalent chromium of cigarette paper detection from this model. The standard uncertainty was 0.12%, the extend uncertainty (α = 0.05, k = 2) was 0.24%.
文章引用:李远根, 王龙, 寇天舒. 评定连续流动法测量烟用接装纸中六价铬的测量不确定度[J]. 分析化学进展, 2017, 7(2): 102-108. https://doi.org/10.12677/AAC.2017.72014

参考文献

[1] 中国合格评定国家认可委员会. CNAS-GL06 化学分析不确定度的评估指南[S]. 2006.
[2] 国家质量监督检验检疫总局发布. JF 1059-1999 测量不确定度评定与表示[S]. 2012.
[3] 中国实验室国家认可委员会. 化学分析中不确定度的评估指南[S].中国计量出版社, 2002.
[4] 庞永强, 陈再根, 陈欢, 等. 卷烟测流烟气中一氧化碳测量不确定度的评定[J]. 现代测量与实验室管理, 2010, 18(1): 30-31.
[5] 张悦. 连续流动法测定水中六价铬不确定度评定[J]. 吉林水利, 2015(5): 50-52.
[6] 王洪波, 赵乐, 郭军伟, 等. 固相萃取净化—连续流动发测定纸中Cr(Ⅵ)[J]. 包装工程, 2016(1): 107-110.
[7] YC/T 60-2015 烟用接装纸六价铬测定连续流动法[S].
[8] YC/T 31-1996 烟草及烟草制品试样的制备和水分测定烘箱法[S].